Influence of a defoaming agent – simethicone – on endoscope cleaning and disinfection: Prospective real-world study

 SFX Search Permissions and Reprints Abstract Background and study aims

Simethicone has been extensively utilized in endoscopy examinations and therapies; however, consensus regarding its impact on endoscopy cleaning is still lacking. The aim of this study was to assess impact of simethicone use during endoscopic examination on efficacy of endoscope cleaning.

Methods

This was a prospective real-world study that involved use of varying concentrations of simethicone in the endoscope biopsy channel and auxiliary water channel.

Results

All simethicone residual amounts and adenosine triphosphate (ATP) values were analyzed every month for 1 year. Use of 1% and 2% concentrations of simethicone generally resulted in variations in residual simethicone levels between the two channels. There was no significant alteration in ATP values in any concentration between the two channels. However, there was a significant difference in ATP values between the two channels at the concentration of 1% simethicone. After 1 year of usage, suspected adherent was observed in the 2% simethicone group, whereas no crystals were detected adhering to the biopsy channel walls in the 1% group or the control group. Sensitivity analysis suggested that the study results did not differ between the gastroscopy and colonoscopy subgroups.

Conclusions

Simethicone may remain in the biopsy and water infusion channels, regardless of whether it is used or not. It is recommended to utilize a simethicone concentration of 1% or less when administering it through the biopsy or auxiliary water channels of the endoscope.

Keywords Quality and logistical aspects - Quality management - Preparation Publication History

Received: 24 November 2024

Accepted after revision: 05 August 2025

Accepted Manuscript online:
11 August 2025

Article published online:
01 September 2025

© 2025. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial-License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/).

Georg Thieme Verlag KG
Oswald-Hesse-Straße 50, 70469 Stuttgart, Germany

Bibliographical Record
Juanjuan Huang, Tingsheng Ling, Junlin Zhang, Lianzhen Wei, Lei Chen, Huiwen Cao, Lei Wang, Yitong Liu, Dongkun Wen, Danrui Ren, Yang Li. Influence of a defoaming agent – simethicone – on endoscope cleaning and disinfection: Prospective real-world study. Endosc Int Open 2025; 13: a26812659.
DOI: 10.1055/a-2681-2659

Comments (0)

No login
gif