According to key facts about deafness and hearing loss, the World Health Organization proposed that “by 2050 nearly 2.5 billion people are projected to have some degree of hearing loss and at least 700 million will require hearing rehabilitation” (https://www.who.int/news-room/fact-sheets/detail/deafness-and-hearing-loss). Hearing technologies for rehabilitation include hearing aids, cochlear implants, and middle-ear implants. Cochlear implantation, among them, has been the major treatment option for severe-to-profound sensorineural hearing loss across all ages from children to the elderly. The prevalence of cochlear implantation is increasing due to factors such as financial support from governments worldwide. Surgical techniques and postoperative care for cochlear implantation have also been optimized in recent years that it can be a day-surgery with minimal invasiveness.
A cochlear implant comprises two main components: internal and external parts. The body of the internal part is embedded in the mastoid bone beneath the skin, and the electrode of the internal part is placed into the cochlea through the round window or cochleostomy during the surgery. The external part mainly comprises a speech processor, a coil, and a microphone. The speech processor and microphone are placed on the auricle of the outer ear, and the coil is positioned behind the ear just on the skin covering the body of the internal part. Communication between the two parts depends on the connection by the magnet inside the coil of the external part. The acoustic signals received by the microphone are processed into digital information by a speech processor and then transmitted through the coil to the electrode. When the cochlear nerve is stimulated by digital information from the electrode, acoustic signals are transferred to the brain, and implanted patients are able to hear again. By inserting the electrode into the cochlea and stimulating the cochlear nerve, cochlear implantation can improve the hearing, speech, and music performance of the hearing-impaired.1,2
The microenvironment of the cochlea is an innate system in nature, and logically, it is disturbed to some extent in response to the insertion of an electrode as a foreign body. These disturbances in the microenvironment due to the insertion certainly have some effect on the performance of the device. It is an interplay between the implant and the circumstances surrounding the electrodes. However, real-time changes inside the cochlea after implantation in terms of electrophysiology at the molecular level can never be explored in vivo in human beings.
Among other tools, the impedance of the electrode is a good and important guide for reflecting the electrophysiological variations inside the cochlea. Because details about the performance of the electrode, magnitude of current flow, and energy demand of the device can be indexed by the detection of impedance,3 field telemetry of impedance has thus become one of the most popular methods of measurements easily accessible for cochlear implantation. Therefore, by observing the variation in impedance, information about intracochlear changes in the surrounding encompassing electrodes can be predicted.3–6 Impedance is also a useful tool for the optimization of most comfortable levels in terms of stimulus and settings for the cochlear implant.7 In addition, impedance can be a potential biomarker highly correlated with imbalance disorder6 and/or performance of hearing as well speech after the surgery.8–11
The first contact between the two main components of a cochlear implant after surgery is called the initial switch-on.12 Subsequently, it is followed by a series of frequency mapping to optimize the performance of a cochlear implant for the commencement of a training program for hearing and speech. Traditionally, initial frequency mapping and training programs can only begin no sooner than 3 weeks postoperatively, regardless of the various surgical approaches currently utilized for cochlear implantation.13,14 Although surgical techniques have been optimized, the initial switch-on cannot be performed earlier than the traditional interval of approximately 1 month postoperatively. Early changes in impedance immediately after insertion were unexplored until recently; however, surgeons worldwide are now trying the initial switch-on earlier than 1 month after implantation.
Experiences about initial switch-on within 24 hours following cochlear implantation have been reported in studies from different countries.8,9,12,15–19 Some of these studies reported that the key issue, among others, for the success of first mapping in <1 day is dedicated at least in part to factors including surgical procedures and wound managements.12,15–18 Details of these factors included but were not limited to microinvasive technique, decreased edema without injury in the peri-incision region, and adverse events to the perilymph following insertion of electrodes.12,15–18 Issues about safety and practicability of earliest switch-on in <1 day postoperatively have been reported previously20; however, serial changes that occur after implantation inside the cochlear using electrophysiology have never been reviewed.
This review discusses the scenario of electrophysiological variation in terms of early switch-on in <1 day postimplantation. Currently, there are three major manufacturers of cochlear implants, and the scenario of changes in the impedance for each manufacturer is different. Arguments on these differences were addressed. We aimed to elucidate the clinical impact of this procedure on implantees.
2. SCENARIO OF CHANGES IN ELECTROPHYSIOLOGICAL PARAMETERS AFTER EARLIEST INITIAL SWITCH-ON WITHIN 24 HOURS OF COCHLEAR IMPLANTATIONElectrically evoked compound action potential (ECAP) remains relatively stable as long as up to 5 to 6 years postimplantation.21 On the contrary, variation in both impedance and ECAP is generally observed days or weeks after the implantation.22,23 These changes are due to factors including cascades of responses to a “foreign body”-like effect resulting from the insertion of electrodes into the inner ear.24,25 Such effects were unclear on the electrophysiological events minutes to hours after the implantation until 2013, when results of earliest initial mapping in <1 day postimplantation were reported for the first time.12 There are indeed differences between electrodes from various manufacturers with respect to scenario of changes in electrophysiological parameters after earliest initial switch-on within 24 hours of cochlear implantation.
3. DEVICES BY COCHLEAR (COCHLEARTM, LANE COVE, NSW, AUSTRALIA) 3.1. Drop in impedance 24 hours after the surgery in comparison to intraoperative monitoringElectrical levels measured after the insertion of an implant into the cochlea of animals tend to drop within days.26 Owing to factors described in the earlier sections about the dilemma of initial switch-on earlier than the traditional scenario of 3 to 4 weeks after cochlear implantation, relevant issues can only be explored in animal models for an extended period in the past. However, an immediate decrease in impedance/ECAP in human beings has been reported within 24 hours after cochlear implantation in comparison to those detected during surgery.12 This phenomenon was later verified in another study from the same group16 and by another team using a different electrode from the same manufacturer.19
Why do impedance and ECAP significantly drop in <1 day postoperatively? Owing to the innate nature of the cochlear system, electrophysiological fluctuation following the surgery must theoretically be related to the reorganization of the milieus around the electrode array on site.27 One possibility could be associated with ups and downs of events provoked by responses to foreign body due to insertion of the implant into the cochlea.28,29 Air bubbles produced due to the implantation, although vaporized soon later, certainly contribute to a surge in impedance intraoperatively.30 Fibers of spiral ganglion could also become swollen shortly after the implantation but recover soon following the invasive effect.31 These responses would trigger the surge in impediment and a quick return for the sensitivity of ganglion neuron, respectively.26 It has also been shown that a sheath-like matrix should have formed surrounding the intracochlear part of the implant immediately after the insertion, which in turn could lead to the rise of the impedance and thus ECAP for contacts of each electrode.4,12,24,32 This matrix of cell cover might comprise materials such as protein adsorption, and immune cells such as macrophages as well as fibroblasts.4 The continuity of these tissue envelope would possibly be disorganized or even damaged due to spurt of cells through transmission of electrical stimulus after initial activation of the device, a divergence effect called “blow-out” in previous studies.4,33,34 Therefore, impedance and ECAP are predicted to be reduced, because logically the integrity of the tissue matrix should be in direct proportion to the magnitude of impedance.33,35 What added to the reduction in impedance would be a hydride layer made after the above events of cell escape, because the layer has been shown to be able to enlarge the area of contact along the array of electrodes.36
It is noteworthy that in the intraoperative monitoring, the values of impedance decreased in an apical-to-basal direction along the cochlea.12 This phenomenon was in line with earlier researches and was again verified later.16,37,38 The underlying mechanism might be associated with the size of electrode array because impedance is inversely proportional to the geometric area of the contact surface.3 Because the contact area increases all the way from apical to the basal array, the basal segment of electrodes would reasonably have the lowest values of impedance, at least for the Nucleus 24RECA implant system.16
3.2. Between 24 hours and 4 weeksWhat follows the drop in 24 hours is an obvious increase in impedance during the first 4 weeks after surgery.16 It is interesting that the behaviors of the trend for segments differed in impedance measurements. A previous study revealed that there is a smooth increase in basal as well as midportion arrays and a minor decrease in the apical array.16 For the apical array, the impedance first increased to the highest level at 1 week before decreasing afterward.16 This could be a special route of evolution exclusive to the earliest switch-on. In some previous studies, evidence had it that the trend of impedance increased within 2 weeks before the initial activation.37,38 In contrast, the impedance values continued to increase from day 1 to 4 weeks after the initial activation for the basal and middle part of the array.16
In addition to the aforementioned events of cell cover and electrical stimulation, fibrosis induced at a later stage also contributed to the end results of the interplay for impedance measurement during this period. Evidences showed that cell cover and fibrosis both lead to increase in impedance, and launched within hours and days postoperatively, respectively.10,27,34,39,40 Effect of electrical stimulation has been revealed to be more obvious on the surface of contact (ie, cell cover) than on the matrix around the electrode (ie, fibrosis), possibly due to structural differences between them.3 Estimates proposed that electrical stimulation could contribute to around 20% decrease in impedance from 2 to 10 weeks, while only around 5% decrease from 1.5 to 5 years postoperatively.3,4,39 Therefore, previous studies showed that cell cover played a much more important role with respect to impedance measurement within the first few weeks after cochlear implantation,3 because it is logical to conjecture that fibrosis is more resistant to electrical stimulation compared with cell cover.10,16 In addition, the basal part of the electrodes has more fibrosis in comparison to the apical part after cochlear implantation.37,41–43 This finding anatomically corresponds to the lateral wall for the basal turn of the cochlea, the area most easily injured on insertion,34,43,44 which in turn could justify the increase in impedance of basal segment during this period. The impedance between 4 and 8 weeks as well as thereafter was stable with little interval differences,16,27,30,34,45 possibly indicating an ultimate equilibrium among factors including cell cover, electrical stimulation, and fibrosis.46
4. DEVICES BY ADVANCED BIONICS (ADVANCED BIONICSTM, STÄFA, SWITZERLAND)Contrary to the above findings, impedance was later observed to be increasing instead of decreasing 1 day after cochlear implantation for the implant system manufactured by a different company (HiFocus 1J; Advanced BionicsTM) for the first time.17 The scenario for the postoperative changes in impedance for various devices (ie, Mid-Scala and Nucleus 24RECA) were also observed to be different.7 Although the results were incongruent with earlier research in animal model, where the reduction in the level of electrical detection persisted days after electrode insertion,26 these studies again verified factors leading to changes in impedance measurements including sheath formation, size of the electrode, fibrosis formation, and electrical stimulation.7,17
However, the mechanisms underlying these differences remain unclear. One possible explanation is the surgical approach. The implant was inserted through the round window in the Advanced BionicsTM researches7,17 and cochleostomy in the aforementioned CochlearTM study.16 According to the animal model, there might be an effect on the cochlea using different techniques, with less tissue formation through the round window approach.47 Another reasonable explanation could be the methodology of monitoring. The timing of impedance measurement for the Advanced BionicsTM device was in the beginning, while for the CochlearTM device, it was measured at the end of the cathodic phase for the biphasic cathodic-anodic pulse, representing “access resistance” and total impedance (ie, sum of access resistance and polarization impedance), respectively.7 This could also be a major factor because access resistance usually increased in a relative smooth way while polarization impedance went down sloping after switch-on.3
There are differences in the device design in terms of electricity. According to Ohm’s law, electrical power is directly proportional to electrical resistance and square of electrical current.17 Basically the electrical current was relatively equal between Advanced BionicsTM and CochlearTM devices.17 However, electrical resistance was much higher in CochlearTM than in Advanced BionicsTM implants, with a maximum of approximately five times of the value.17 This in turn could lead to a much bigger pulse of electrical power passing by way of electrodes to the cell cover and fibrosis surrounding the CochlearTM devices and therefore a greater effect of “blow-out” than in the Advanced BionicsTM devices.17 The authors also argued about the possibility of sheath formation effect,4,48 which might result from increased quantity of cell cover and fibrosis formation due to different positions inside the cochlear space for devices from different manufacturers.49,50 Although this effect could lead to an increased level of impedance, it remains to be investigated.
For the intraoperative measurements, values of impedance telemetry were increased in the basal-to-apical direction of the cochlea,7 which was in line with a previous study that used device from another manufacturer.16 For the Advanced BionicsTM device, the contact size was also decreased in the basal-to-apical direction,7 and same was observed in CochlearTM device.16 Thus, this tendency can be expected from the trend of contact size, because impedance is inversely proportional to the contact area of electrodes.3 The scenario of impedance from 1 day to 2 months was relatively smooth without significant changes.7,17
5. DEVICES BY MED-EL (MED-ELTM, INNSBRUCK, TYROL, AUSTRIA)The impedances of the Med-ElTM device were revealed to be significantly decreased when measured intraoperatively in comparison to those measured 1 day later.8,9,51 The trend was compatible with that noted in devices of Advanced BionicsTM7,17 and different from that of CochlearTM devices.12,15,16 There was no significant change even after 1 year postoperatively.9
On the contrary, the impedances of the Med-ElTM device were revealed to be significantly increased when measured intraoperatively in comparison to those measured in the recovery room on the same day under conscious sedation (CS).8 The authors ascribed the reasons to be air bubbles or inflammatory process.8 There was no significant change from 2 to 8 weeks postoperatively.8 However, the limit of the study resided in the fact that the sample size was small (ie, nine implantees in total).8 Further studies are necessary to clarify the scenario of impedance change after early switch-on under CS.
In conclusion, hearing loss is a major disease with social burden that is progressively increasing in recent years. Overthe past few decades, cochlear implantation has become a well-developed option for the optimal rehabilitation of patients with severe-to-profound hearing impairment. As a foreign body, the inserted implant certainly has some effect on the innate circumstances of electrophysiology inside the cochlea. Because the real conditions at the molecular level can never be explored in vivo with respect to human beings, impedance should be a good indicator for monitoring changes in the microenvironment of the cochlea. Although findings from animal models suggest that electrophysiological parameters begin to fluctuate soon after insertion, early changes in impedance after cochlear implantation in human beings remain unclear because traditionally the initial measurement can only be made approximately 1 month after surgery. The experiences of earliest initial switch-on from various teams worldwide can now be found in the relevant literature database. Evidence has revealed that the final results of the measured values for impedance depend on the equilibrium among parameters including cell cover formation, growth of fibrosis, electrode design, and electrical stimulation. Further studies are needed to address the association between impedance values and parameters of clinical entities, which in turn could promise the development of a reliable biomarker for a better performance of cochlear implantation.
REFERENCES 1. Chen JK, Chuang AY, McMahon C, Hsieh JC, Tung TH, Li LP. Music training improves pitch perception in prelingually deafened children with cochlear implants. Pediatrics. 2010;125:e793–800. 2. Chen JK, Chuang AY, McMahon C, Tung TH, Li LP. Contribution of nonimplanted ear to pitch perception for prelingually deafened cochlear implant recipients. Otol Neurotol. 2014;35:1409–14. 3. Tykocinski M, Cohen LT, Cowan RS. Measurement and analysis of access resistance and polarization impedance in cochlear implant recipients. Otol Neurotol. 2005;26:948–56. 4. Newbold C, Richardson R, Huang CQ, Milojevic D, Cowan R, Shepherd R. An in vitro model for investigating impedance changes with cell growth and electrical stimulation: implications for cochlear implants. J Neural Eng. 2004;1:218–27. 5. Newbold C, Farrington A, Peters L, Cowan R, Needham K. Electropermeabilization of adherent cells with cochlear implant electrical stimulation in vitro. Audiol Neurootol. 2014;19:283–92. 6. Li LP, Wei JJ, Chen JK, Wu DC, Coelho DH. Early elevation and normalization of electrode impedance in patients with enlarged vestibular aqueduct undergoing cochlear implantation. Otol Neurotol. 2022;43:e535–9. 7. Wei JJ, Tung TH, Li LP. Evolution of impedance values in cochlear implant patients after early switch-on. PLoS One. 2021;16:e0246545. 8. Aldhafeeri AM, Saleh S, Almuhawas F, Hagr A. Feasibility of day surgery for cochlear implantation under conscious sedation with same-day fitting. J Int Adv Otol. 2020;16:303–8. 9. Alhabib SF, Abdelsamad Y, Yousef M, Alzhrani F, Hagr A. Effect of early activation of cochlear implant on electrode impedance in pediatric population. Int J Pediatr Otorhinolaryngol. 2021;140:110543. 10. Alsabellha RM, Hagr A, Al-Momani MO, Garadat SN. Cochlear implant device activation and programming: 5 days postimplantation. Otol Neurotol. 2014;35:e130–4. 11. Hagr A, Garadat SN, Al-Momani M, Alsabellha RM, Almuhawas FA. Feasibility of one-day activation in cochlear implant recipients. Int J Audiol. 2015;54:323–8. 12. Chen JK, Chuang AY, Sprinzl GM, Tung TH, Li LP. Impedance and electrically evoked compound action potential (ECAP) drop within 24 hours after cochlear implantation. PLoS One. 2013;8:e71929. 13. Mann WJ, Gosepath J. Technical note: minimal access surgery for cochlear implantation with MED-EL devices. ORL J Otorhinolaryngol Relat Spec. 2006;68:270–2. 14. Papsin BC, Gordon KA. Cochlear implants for children with severe-to-profound hearing loss. N Engl J Med. 2007;357:2380–7. 15. Chen JK, Chuang AY, Sprinzl GM, Tung TH, Li LP. Safety and feasibility of initial frequency mapping within 24 hours after cochlear implantation. Acta Otolaryngol. 2015;135:592–7. 16. Hu HC, Chen JK, Tsai CM, Chen HY, Tung TH, Li LP. Evolution of impedance field telemetry after one day of activation in cochlear implant recipients. PLoS One. 2017;12:e0173367. 17. Lin DP, Chen JK, Tung TH, Li LP. Differences in the impedance of cochlear implant devices within 24 hours of their implantation. PLoS One. 2019;14:e0222711. 18. Chen J, Chen B, Shi Y, Li Y. A retrospective review of cochlear implant revision surgery: a 24-year experience in China. Eur Arch Otorhinolaryngol. 2021;279:1211–20. 19. Sunwoo W, Jeon HW, Choi BY. Effect of initial switch-on within 24 hours of cochlear implantation using slim modiolar electrodes. Sci Rep. 2021;11:22809. 20. Coelho DH, Shiao AS, Li LP. Very early activation of cochlear implants: a review of the literature. J Chin Med Assoc. 2023;86:7–10. 21. Brown CJ, Abbas PJ, Etlert CP, O’Brient S, Oleson JJ. Effects of long-term use of a cochlear implant on the electrically evoked compound action potential. J Am Acad Audiol. 2010;21:5–15. 22. Henkin Y, Kaplan-Neeman R, Muchnik C, Kronenberg J, Hildesheimer M. Changes over time in electrical stimulation levels and electrode impedance values in children using the Nucleus 24M cochlear implant. Int J Pediatr Otorhinolaryngol. 2003;67:873–80. 23. Henkin Y, Kaplan-Neeman R, Kronenberg J, Migirov L, Hildesheimer M, Muchnik C. A longitudinal study of electrical stimulation levels and electrode impedance in children using the Clarion cochlear implant. Acta Otolaryngol. 2006;126:581–6. 24. Kawano A, Seldon HL, Clark GM, Ramsden RT, Raine CH. Intracochlear factors contributing to psychophysical percepts following cochlear implantation. Acta Otolaryngol. 1998;118:313–26. 25. Nadol JB Jr, Eddington DK, Burgess BJ. Foreign body or hypersensitivity granuloma of the inner ear after cochlear implantation: one possible cause of a soft failure? Otol Neurotol. 2008;29:1076–84. 26. Su GL, Colesa DJ, Pfingst BE. Effects of deafening and cochlear implantation procedures on postimplantation psychophysical electrical detection thresholds. Hear Res. 2008;241:64–72. 27. Marsella P, Scorpecci A, Pacifico C, Resca A, Vallarino MV, Ingrosso A, et al. Safety and functional results of early cochlear implant switch-on in children. Otol Neurotol. 2014;35:277–82. 28. Dodson HC. Loss and survival of spiral ganglion neurons in the guinea pig after intracochlear perfusion with aminoglycosides. J Neurocytol. 1997;26:541–56. 29. Leake-Jones PA, O’Reilly BF, Vivion MC. Neomycin ototoxicity: ultrastructural surface pathology of the organ of Corti. Scan Electron Microsc. 1980;3:427–34. 30. Wolf-Magele A, Schnabl J, Edlinger S, Pok SM, Schoerg P, Sprinzl GM. Postoperative changes in telemetry measurements after cochlear implantation and its impact on early activation. Clin Otolaryngol. 2015;40:527–34. 31. Juiz JM, Rueda J, Merchan JA. Reversible damage to the nerve fibres in the organ of Corti after surgical opening of the cochlea in the rat. Acta Otolaryngol. 1988;106:29–33. 32. Agrawal V, Newbold C. Computer modelling of the cochlea and the cochlear implant: a review. Cochlear Implants Int. 2012;13:113–23. 33. Charlet de Sauvage R, Lima da Costa D, Erre JP, Aran JM. Electrical and physiological changes during short-term and chronic electrical stimulation of the normal cochlea. Hear Res. 1997;110:119–34. 34. Hughes ML, Vander Werff KR, Brown CJ, Abbas PJ, Kelsay DM, Teagle HF, et al. A longitudinal study of electrode impedance, the electrically evoked compound action potential, and behavioral measures in nucleus 24 cochlear implant users. Ear Hear. 2001;22:471–86. 35. Clark GM, Shute SA, Shepherd RK, Carter TD. Cochlear implantation: osteoneogenesis, electrode-tissue impedance, and residual hearing. Ann Otol Rhinol Laryngol Suppl. 1995;166:40–2. 36. Brummer SB, Turner MJ. Electrochemical considerations for safe electrical stimulation of the nervous system with platinum electrodes. IEEE Trans Biomed Eng. 1977;24:59–63. 37. Busby PA, Plant KL, Whitford LA. Electrode impedance in adults and children using the Nucleus 24 cochlear implant system. Cochlear Implants Int. 2002;3:87–103. 38. Paasche G, Bockel F, Tasche C, Lesinski-Schiedat A, Lenarz T. Changes of postoperative impedances in cochlear implant patients: the short-term effects of modified electrode surfaces and intracochlear corticosteroids. Otol Neurotol. 2006;27:639–47. 39. Newbold C, Mergen S, Richardson R, Seligman P, Millard R, Cowan R, et al. Impedance changes in chronically implanted and stimulated cochlear implant electrodes. Cochlear Implants Int. 2014;15:191–9. 40. Stadelmann WK, Digenis AG, Tobin GR. Physiology and healing dynamics of chronic cutaneous wounds. Am J Surg. 1998;176(2A Suppl):26S–38S. 41. Nadol JB Jr, Shiao JY, Burgess BJ, Ketten DR, Eddington DK, Gantz BJ, et al. Histopathology of cochlear implants in humans. Ann Otol Rhinol Laryngol. 2001;110:883–91. 42. Zadrozniak M, Szymanski M, Siwiec H, Broda T. Impedance changes in cochlear implant users. Otolaryngol Pol. 2011;65:214–7. 43. Somdas MA, Li PM, Whiten DM, Eddington DK, Nadol JB Jr. Quantitative evaluation of new bone and fibrous tissue in the cochlea following cochlear implantation in the human. Audiol Neurootol. 2007;12:277–84. 44. Rebscher SJ, Hetherington A, Bonham B, Wardrop P, Whinney D, Leake PA. Considerations for design of future cochlear implant electrode arrays: electrode array stiffness, size, and depth of insertion. J Rehabil Res Dev. 2008;45:731–47. 45. Molisz A, Zarowski A, Vermeiren A, Theunen T, De Coninck L, Siebert J, et al. Postimplantation changes of electrophysiological parameters in patients with cochlear implants. Audiol Neurootol. 2015;20:222–8. 46. Joseph J, Mohanty M, Mohanan PV. Role of immune cells and inflammatory cytokines in regulation of fibrosis around silicone expander implants. J Mater Sci Mater Med. 2010;21:1665–76. 47. Burghard A, Lenarz T, Kral A, Paasche G. Insertion site and sealing technique affect residual hearing and tissue formation after cochlear implantation. Hear Res. 2014;312:21–7. 48. Ishai R, Herrmann BS, Nadol JB Jr, Quesnel AM. The pattern and degree of capsular fibrous sheaths surrounding cochlear electrode arrays. Hear Res. 2017;348:44–53. 49. Gazibegovic D, Bero EM. Multicenter surgical experience evaluation on the Mid-Scala electrode and insertion tools. Eur Arch Otorhinolaryngol. 2017;274:1147–51. 50. Dhanasingh A, Jolly C. An overview of cochlear implant electrode array designs. Hear Res. 2017;356:93–103. 51. Hu HC, Chen JK, Li LP, Chen WK, Huang MJ, Yeh CH, et al. Evolution of impedance values in two different electrode array designs following activation of cochlear implants 1 day after surgery: a study of 58 patients. Clin Otolaryngol. 2020;45:584–90.